Clarity - Getting Started First “Analysis”
43
the retention time, the quantity will be zero. If
the calibration compound is found and
allocated, but no calibration curve exists, the
Peak Type column will display the word Error.
Caution! The chromatogram typically uses calibration
curves from the calibration file whose name is
shown in the right-hand section of Results tab of
the Chromatogram window. Therefore, this is
calibration by reference. Its advantage lies in
easy recalibration and changes in case of use
of a calibrated file for more chromatograms.
Nevertheless, all calibration curves are simul-
taneously saved directly in the chromatogram,
including their history. Results according to
these saved calibration curves (including then
valid chromatogram modifications) can be
recalled if you select the relevant method in the
dialog for chromatogram selection - Open
Chromatogram in the Method field.
5.7 Printing
Comments
To facilitate the printing, all the important
windows contain icons
and .
The report style file actually defines the layout
and contents of the printed report.
Information is divided into subsections, corre-
sponding to the displayed tabs. On each tab
you can select a more detailed arrangement.
To include the entire section in the report,
check the Print checkbox in the top-left corner
of the tab. The selected section is marked with
a tick on the tab label.
TIP: A tab can also be selected by double-clicking the
left mouse button on the tab.
By default there are different report styles
predefined according to the dialog from which