Clarity - Getting Started Installation
12
2.3.1 Standard cable for Clarity station
Signal cables
Labelled “DET1” to “DET4” (according to the
number of channels), the cables are supplied
as standard without connectors with only
stripped, tinned endings – red (+), white (-) and
shielding (analogue ground).
Starting (marker) cables
Labelled IN1” to “IN4” (according to the
number of channels), ended with CINCH
connector. One cable with free leads [red (+),
shielding (digital ground)] for connection di-
rectly to the chromatograph or valve and one
cable ended with a button for cases where
starting contact is not available and it is nec-
essary to perform start manually are supplied
for each starting cable.
Cables of the digital outputs
Relay contacts labelled, "OUT 1R" to "OUT 4R"
(according to the number of channels), are
ended with free leads. They are used for
synchronizing autosamplers in the active
sequence without an AS Control module.
At the converter end, the cable is always
ended with CANNON SUB D 37-pin connector.
2.3.2 Chromatograph
Connect cables according to one of the
following diagrams in Fig. 1. Use symmetrical
connection only in the case that you are sure
that the chromatograph/detector is equipped
with symmetrical output – it is necessary to
read through the instructions for the
corresponding chromatograph.