Clarity - Getting Started Installation
13
Starting cable
Signal cable
START (REMOTE, …)
Red
Shielding
SYMMETRICAL CONNECTION
+ (+1V, Hi, …)
- (-1V, Lo, …)
COM (0, GND)
Red
Shielding
White
Digital output (Relay)
START (REMOTE, …)
1V (100mV, Hi, +, …)
0 (COM, AGND, …)
NONSYMMETRICAL CONNECTION
Red
Shielding
Red
Shielding
White
Starting cable
Signal cable
Digital output (Relay)
Fig. 1. Connection of Clarity with chromatograph
All current DataApex A/D Converters INT7,
INT9, U-PAD, U-PAD2 and Net-PAD use the same
standard INT7 Connector.
Note: The INT5 cable (obsolete type) requires a
reduction to the INT7 card, which can be ordered
separately.
Detailed description of individual connectors
wiring is provided in the Reference Guide in the
chapter Technical Specification.
Principles of connection of signal cables:
Signal inputs of the A/D converters are
symmetrical: + (red), - (white), analogue earth
(copper braiding).
Caution! Shielding must be connected. It serves not only
as the shielding, but also as the analogue earth
against which measurement takes place. In the
case of asymmetrical output of a detector (only
two connectors) shielding must be connected to
the white lead! No lead of the signal cable may
remain unconnected.
Try to connect to the output of the
chromatograph detector with the largest
possible level of signal, usually indicated as
INTEGRATOR (signal approx. 1V). The level of
the signal on the output marked RECORDER
is usually only approx. 10mV.