Clarity INT9 A/D Guide de l'utilisateur Page 38

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 60
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 37
Clarity - Getting Started First “Analysis”
38
5.5.2 Chromatogram processing
Comments
After stopping an analysis it is advisable to
check and optimize the integration parameters.
The key parameters are Peak Width
and Threshold. Their simplified definition is:
The Peak Width parameter determines that all
distinctly narrow peaks, compared to the de-
fined value, will not be integrated.
The Threshold parameter determines the noise
threshold, which means that only peaks of at
least double the height of the defined value are
detected.
Note: In reality, however, the situation is more complex
as it always combines the mutual effects of both
parameters and the peak shape.
Fig. 16. Method Setup - Integration
Both parameters can be applied to the whole
chromatogram (Global) or to a defined interval
(Local).
The
Global (Local) Peak Width and
Global (Local) Threshold interactive
functions in the
Chromatogram window will help
Vue de la page 37