Clarity INT9 A/D Spécifications Page 29

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 35
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 28
Clarity - INT9 A/D converter Troubleshooting
29
In the Models list select the DataApex
Acquisition device, Model 9.00 and press the
Next button.
In the remaining dialogs press the Next
button.
Restart the PC.
Re-check the driver in the
Device Manager as
desribed in chapter 4.3.3.
4.4 Data Acquisition – non-functional
A gray icon, label “Disabled”, and Data Acquisition command not functional
Other manifestations of this error are also:
Method Setup - Acquisition tab missing, Method –
Acquisition command non-functional, Run,
Stop, Abort commands non-functional in the
windows
Single Analysis and Sequence).
Possible causes are:
a) Detector not allocated to instrument:
Open the System Configuration dialog from the
main
Clarity window using the System -
Configuration command and check the tab of
the corresponding instrument -
Instrument x. If it
has no allocated detectors, add them.
Solution: In the left-hand list Setup Control Modules
select the correct detector that is connected to
the A/D card you are using and link it to the
corresponding instrument on the right.
Vue de la page 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 35

Commentaires sur ces manuels

Pas de commentaire